Bsv - Keysight Technologies E5260 Series Programming Manual

Parametric measurement solutions
Hide thumbs Also See for E5260 Series:
Table of Contents

Advertisement

Syntax
Parameters
Example
Statements

BSV

The BSV command specifies and sets the voltage search source for the binary
search measurement. This command setting is cleared by the BSI command. After
search stops, the search channel forces the value specified by the BSM command.
BSV chnum,range,start,stop[,Icomp]
chnum :
Search source channel number. The value must be slot number where
the module has been installed. Integer expression. 1 to 8 (1 and 5 are
not available for HPSMU). See
range :
Output ranging type. Integer expression. The output range will be set to
the minimum range that covers both start and stop values. For the
limited auto ranging, the instrument never uses the range less than the
specified range. See
start, stop : Search start or stop voltage (in V). Numeric expression. See
on page
4-14. The start and stop parameters must have different values.
0 to ±100, or 0 to ±200 for E5280B/E5290A
Icomp :
Current compliance value (in A). Numeric expression. See
page
4-14. If you do not specify Icomp, the previous value is set. Zero
amps (0 A) is not allowed for Icomp.
OUTPUT @E5270;"BSV 1,0,0,20,1E-6"
Keysight E5260/E5270 Programming Guide, Edition 4
Table 4-1 on page
Table 4-4 on page
4-13.
Command Reference
Command Reference
4-9.
Table 4-6
Table 4-6 on
4- 39

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

E5270 seriesE5260aE5262aE5263aE5270b

Table of Contents