The BSV command specifies and sets the voltage search source for the binary
search measurement. This command setting is cleared by the BSI command. After
search stops, the search channel forces the value specified by the BSM command.
BSV chnum,range,start,stop[,Icomp]
chnum :
Search source channel number. The value must be slot number where
the module has been installed. Integer expression. 1 to 8 (1 and 5 are
not available for HPSMU). See
range :
Output ranging type. Integer expression. The output range will be set to
the minimum range that covers both start and stop values. For the
limited auto ranging, the instrument never uses the range less than the
specified range. See
start, stop : Search start or stop voltage (in V). Numeric expression. See
on page
4-14. The start and stop parameters must have different values.
0 to ±100, or 0 to ±200 for E5280B/E5290A
Icomp :
Current compliance value (in A). Numeric expression. See
page
4-14. If you do not specify Icomp, the previous value is set. Zero
amps (0 A) is not allowed for Icomp.
OUTPUT @E5270;"BSV 1,0,0,20,1E-6"