Syntax
Parameters
Example
Statements
LGV
The LGV command specifies the voltage monitor channel and its search parameters
for the linear search measurement.
This command ignores the RV command setting.
This command setting is cleared by the LGI command.
LGV chnum,mode,range,target
chnum :
Search monitor channel number. The value must be slot number where
the module has been installed. Integer expression. 1 to 8 (1 and 5 are
not available for HPSMU). See
mode :
Search mode. Integer expression.
0 : If the measured value ≤ target, it is the search result data.
1 : If the measured value ≥ target, it is the search result data.
range :
Measurement ranging type. Integer expression. The measurement range
will be set to the minimum range that covers the target value. For the
limited auto ranging, the instrument never uses the range less than the
specified range. See
target :
Search target voltage (in V). Numeric expression.
0 to ±100, or 0 to ±200 for E5280B/E5290A
OUTPUT @E5270;"LGV 1,2,12,3"
Keysight E5260/E5270 Programming Guide, Edition 4
Table 4-1 on page
Table 4-2 on page
4-10.
Command Reference
Command Reference
4-9.
4- 65