Ers - Keysight Technologies E5260 Series Programming Manual

Parametric measurement solutions
Hide thumbs Also See for E5260 Series:
Table of Contents

Advertisement

Syntax
Query Response
Example
Statements
See Also

ERS?

The ERS? command returns the status of the digital I/O port (16 paths).
ERS?
pattern <CR/LF^EOI>
pattern returns the decimal value of the port status.
The status of each port is designated by 0 or 1 that has the following meaning:
0: TTL high level (approx. 2.4 V)
1: TTL low level (approx. 0.8 V)
OUTPUT @E5270;"ERS?"
ENTER @E5270;A
PRINT "Port Status=";A
For example, 255 (0000000011111111) is returned when the port 0 to 7 have been
set to the TTL low level and the port 8 to 15 have been set to the TTL high level.