WM
The WM command enables or disables the automatic abort function for the staircase
sweep sources and the pulsed sweep source. The automatic abort function stops the
measurement when one of the following conditions occurs:
Compliance on the measurement channel
Compliance on the non-measurement channel
Overflow on the AD converter
Oscillation on any channel
This command also sets the post measurement condition for the sweep sources.
After the measurement is normally completed, the staircase sweep sources force the
value specified by the post parameter, and the pulsed sweep source forces the pulse
base value.
If the measurement is stopped by the automatic abort function, the staircase sweep
sources force the start value, and the pulsed sweep source forces the pulse base
value after sweep.
WM abort[,post]
abort :
Automatic abort function. Integer expression.
1: Disables the function. Initial setting.
2: Enables the function.
post :
Source output value after the measurement is normally completed.
Integer expression.
1: Start value. Initial setting.
2: Stop value.
If this parameter is not specified, the sweep sources force the start
value.
The E5260/E5270 returns the data measured before any abort condition is detected.
Dummy data 199.999E+99 will be returned for the data after abort.
OUTPUT @E5270;"WM 2"
OUTPUT @E5270;"WM 2,2"