Syntax
Parameters
Example
Statements
BSM
The BSM command specifies the search source control mode in the binary search
measurement (MM15), and enables or disables the automatic abort function. The
automatic abort function stops the measurement when one of the following
conditions occurs:
Compliance on the measurement channel
Compliance on the non-measurement channel
Overflow on the AD converter
Oscillation on any channel
This command also sets the post search condition for the binary search sources.
After the search measurement is normally completed, the binary search sources
force the value specified by the post parameter.
If the search operation is stopped by the automatic abort function, the binary search
sources force the start value after search.
BSM mode
abort
,
[,
mode :
Source output control mode, 0 (normal mode) or 1 (cautious mode).
If you do not enter this command, the normal mode is set. See
4-1.
abort :
Automatic abort function. Integer expression.
1: Disables the function. Initial setting.
2: Enables the function.
post :
Source output value after the measurement is normally completed.
Integer expression.
1: Start value. Initial setting.
2: Stop value.
3: Output value when the search target value is get.
If you do not specify post, the search source forces the start value.
OUTPUT @E5270;"BSM 1,2,3"