Figure 2-19
Keysight E5270
NOTE
Remote Mode Functions
Trigger Function
Trigger Function
The Keysight E5260/E5270 can be synchronized with other equipment, such as
capacitance meters, voltmeters, ammeters, probers, handlers and so on, by using the
following terminals:
•
Ext Trig In
BNC connector. Only for trigger input (to receive trigger).
•
Ext Trig Out
BNC connector. Only for trigger output (to send trigger).
•
Digital I/O
D-Sub 25-pin connector. Sixteen paths are available for the trigger port. Each
path can be used for either input or output. For the pin assignment and
accessories, refer to
Figure 2-19
shows a connection example of the E5260/E5270 and another device.
Connecting Trigger Input/Output
In
Other device
In
Ext Trig
To use the digital I/O port for the trigger input/output port, send the TGP command.
DIO 1 to DIO 16 can be used for the trigger input/output port. See
2- 30
Keysight E5260/E5270 Programming Guide, Edition 4
"Digital I/O
Port".
Out
13
Out
out
in
Example connection
out
in
1
25
14
Digital I/O
Table
2-2.