NOTE
To Use
Synchronous
Sweep Source
Figure 2-9
7. After the sweep measurement, the pulsed source forces the pulse base value, and
the sweep source forces the start or stop value, as specified by the WM
command, and keeps it.
For 0 V output, enter the DZ command that is used to memorize the present
settings of the channel and change the output to 0 V.
The WM command sets the automatic abort function and the output after
measurement. The WV command sets the sweep voltage, and the WI command sets
the sweep current. The start and stop values must have the same polarity for log
sweep.
The PT command sets the pulse timing parameters, such as pulse width and pulse
period. The PV command sets the voltage pulse, and the PI command sets current
pulse.
One more channel can be set up as a sweep source that has the output synchronized
with the staircase sweep. Refer to
After the measurement, the synchronous sweep source forces the start or stop value,
as specified by the WM command, and keeps it.
Synchronous Sweep
Voltage or current
Stop value
WT/WM/WV/WI
Trigger (e.g. XE)
Previous value
Start value
Voltage or current
Stop value
WSV/WSI
Previous value
Start value
Keysight E5260/E5270 Programming Guide, Edition 4
Remote Mode Functions
"Synchronous Output" on page
Measurement Modes
2-21.
Primary sweep
Synchronous sweep
Time
2- 13