Table 3-3
Sub perform_meas(ByVal session As IMessage, ByVal t() As Integer)
Dim i As Integer = 0
Dim j As Integer = 0
Dim nop1 As Integer = 1
Dim nop2 As Integer = 1
Dim data(nop2 - 1, nop1 - 1) As String
Dim value As String = "Id (uA), Time (sec), Status"
Dim fname As String = "C:\Keysight\prog_ex\data2.txt"
Dim title As String = "Measurement Result"
Dim msg As String = "No error."
Dim err As Integer = 0
Dim vd As Double = 3
Dim vg As Double = 1
Dim idcomp As Double = 0.05
Dim igcomp As Double = 0.01
Dim orng As Integer = 0
Dim mrng As Integer = 0
session.WriteString("FMT 1" & vbLf)
session.WriteString("TSC 1" & vbLf)
session.WriteString("FL 0" & vbLf)
session.WriteString("AV 10,1" & vbLf)
session.WriteString("DV " & t(3) & ",0,0,0.1" & vbLf)
session.WriteString("DV " & t(2) & ",0,0,0.1" & vbLf)
session.WriteString("DV " & t(1) & "," & orng & "," & vg & "," & igcomp & vbLf)
session.WriteString("DV " & t(0) & "," & orng & "," & vd & "," & idcomp & vbLf)
session.WriteString("MM 1," & t(0) & vbLf)
session.WriteString("CMM " & t(0) & ",1" & vbLf)
session.WriteString("RI " & t(0) & "," & mrng & vbLf)
session.WriteString("ERR? 1" & vbLf) : err = session.ReadString(4 + 2)
If err <> 0 Then session.WriteString("DZ" & vbLf) : GoTo Check_err
Line
2 to 11
Declares variables used through the project. And sets the proper values.
13 to 18
Declares variables and sets the value.
19 to 22
Sets the data output format, time stamp data output mode, SMU filter, and averaging.
23 to 26
Applies voltage to device.
27 to 29
Sets the measurement mode, channel measurement mode, and measurement range.
30 to 31
Checks if an error occurred. If an error is detected, forces 0 V and goes to Check_err.
A program example of a spot measurement is shown below. This example measures
MOSFET drain current.
Spot Measurement Example
Keysight E5260/E5270 Programming Guide, Edition 4
't(0): Drain
't(1): Gate
't(2): Source
't(3): Substrate
'enables time stamp output
'sets filter off
'sets number of samples for 1 data
'out= 0 V, comp= 0.1 A
'out= 0 V, comp= 0.1 A
'1: spot measurement
'1: current measurement
Description
Programming Examples
Spot Measurements
'1
'13
'19
'29
3- 13