Dfm - Keysight Technologies E5260 Series Programming Manual

Parametric measurement solutions
Hide thumbs Also See for E5260 Series:
Table of Contents

Advertisement

Syntax
Parameters
Example
Statements

DFM

The DFM command selects the data display format on the front panel LCD.
The *RST command or the device clear selects scientific.
DFM format
format :
Data display format. Integer expression.
format
0
1
OUTPUT @E5270;"DFM 0"
Keysight E5260/E5270 Programming Guide, Edition 4
Description
Engineering.
+/− sign, 6 digits numeric value with arithmetic point,
and unit.
Example: +123.456mA
Scientific.
+/− sign, 4 digits numeric value with arithmetic point,
exponential part (E, +/− sign, and 1 or 2 digits numeric
value), and unit.
Example: +1.234E-1A
Command Reference
Command Reference
4- 47