Syntax
Query Response
Example
Statements
Syntax
Query Response
Example
Statements
NUB?
The NUB? query command checks the number of measurement data in the output
data buffer, and stores the results in the output data buffer (query buffer).
NUB?
Number of measurement data<CR/LF^EOI>
OUTPUT @E5270;"NUB?"
ENTER @E5270;A
*OPC?
The *OPC? command monitors the pending operations, and places ASCII character
1 into the output queue when all pending operations are completed. Also this
command sets/clears the operation complete (OPC) bit in the standard event status
register as follows:
If there are no pending operations, sets the OPC bit to 1.
If there are any pending operations, sets the OPC bit to 0.
The bit will be set to 1 when all pending operations are completed.
*OPC?
1<CR/LF^EOI>
No response will be returned until all pending operations are completed.
OUTPUT @E5270;"*OPC?"
ENTER @E5270;A
Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
4- 85