The PT command sets the hold time, pulse width, and pulse period. The WM
command sets the automatic abort function and the output after measurement. The
PWV sets the pulsed sweep voltage, and the PWI sets the pulsed sweep current. The
start, stop, and pulse base values must have the same polarity for log sweep.
One more channel can be set up as a staircase sweep source that has the output
synchronized with the pulsed sweep. Refer to
After the measurement, the synchronous sweep source forces the start value, and
keeps it.
Synchronous Sweep
Voltage or current
Stop value
PT/WM/PWV/PWI
Trigger (e.g. XE)
Start value
Previous value
Pulse base value
Voltage or current
Stop value
WSV/WSI
Previous value
Start value
The WSV command sets the sweep voltage, and the WSI command sets the sweep
current. You can use the same output mode (voltage or current) as the pulsed sweep.
The start and stop values must have the same polarity for log sweep.
An automatic abort function is available. Refer to
page
2-23.
Even if the automatic abort function is disabled, the E5260/E5270 automatically
stops measurement if power compliance is enabled for the sweep source and the
power compliance or an automatic abort condition is detected.
Keysight E5260/E5270 Programming Guide, Edition 4
Remote Mode Functions
Measurement Modes
"Synchronous Output" on page
Pulsed sweep
Synchronous sweep
"Automatic abort Function" on
2-21.
Time
2- 11