FL
This command sets the connection mode of a filter for each channel.
A filter is mounted on each module. It assures clean source output with no spikes or
overshooting.
FL mode[,chnum[,chnum...[,chnum]...]]
A maximum of eight channels can be set.
mode :
Status of the filter. Integer expression.
0: Disconnect (initial setting).
1: Connect.
chnum :
Channel number. The value must be slot number where the module has
been installed. Integer expression. 1 to 8 (1 and 5 are not available for
HPSMU). See
If you do not specify chnum, the FL command sets the same mode for all channels.
OUTPUT @E5270;"FL"
OUTPUT @E5270;"FL 0,1,3,5"