Download Print this page
Keysight Technologies 5061B Manual

Keysight Technologies 5061B Manual

Security feature

Advertisement

Quick Links

E5061B/E5063A Security Feature
Rev. 6.0
March 2018
Copyright 2010-2018 Keysight Technologies
Page 1 of 16

Advertisement

loading

Summary of Contents for Keysight Technologies 5061B

  • Page 1 E5061B/E5063A Security Feature Rev. 6.0 March 2018 Copyright 2010-2018 Keysight Technologies Page 1 of 16...
  • Page 2 Contacting Keysight Sales and Service Offices Assistance with test and measurements needs and information on finding local Keysight offices are available on the internet at, http://www.keysight.com/find/assist. If you do not have access to the internet, please contact your field engineer. Note: In any correspondence or telephone conversations, refer to the signal generator by its model number and full serial number.
  • Page 3 Product Declassification and Security Product Name: Network Analyzer Model Number(s): E5061B/E5063A Product Declassification and Security Model Number(s): E5061B/E5063A Product Name: Network Analyzer Product Family Name: ENA This document describes instrument security features and the steps to declassify an instrument through memory sanitization or removal. Table of Contents Terms and Definitions.
  • Page 4: Terms And Definitions

    Product Declassification and Security Product Name: Network Analyzer Model Number(s): E5061B/E5063A Terms and Definitions Definitions: Clearing – Clearing is the process of eradicating the data on media before reusing the media so that the data can no longer be retrieved using the standard interfaces on the instrument. Clearing is typically used when the instrument is to remain in an environment with an acceptable level of protection.
  • Page 5: Instrument Memory

    Product Declassification and Security Product Name: Network Analyzer Model Number(s): E5061B/E5063A Instrument Memory This section contains information on the types of memory available in your instrument. It explains the size of memory, how it is used, its location, volatility, and the sanitization procedure. Summary of instrument memory - base instrument Location in Purpose/Cont...
  • Page 6 Product Declassification and Security Product Name: Network Analyzer Model Number(s): E5061B/E5063A 150 GB – [Revision Prefix: YD488 and above] E5063A (Solid State Drive) 80 GB – [Revision Prefix: KA, KD, KE] 150 GB – [Revision Prefix: KE, Memory for DSP Data Measurement A51/ A71 DSP...
  • Page 7 Product Declassification and Security Product Name: Network Analyzer Model Number(s): E5061B/E5063A and above, A1 Analog Module (Opt 2D5/ 2H5) A3 Analog Module (Opt 205/215/235/245/2 65/285) Page 7 of 16...
  • Page 8: Memory Clearing, Sanitization And/Or Removal

    Product Declassification and Security Product Name: Network Analyzer Model Number(s): E5061B/E5063A Memory Clearing, Sanitization and/or Removal Procedures This section explains how to clear, sanitize, and remove memory from your instrument for all memory types. <Memory type> Description and Main Memory for Windows Operating system memory purpose Size Windows XP: 2 GB, Windows 7: 4 GB, Windows 10: 4 GB...
  • Page 9 Product Declassification and Security Product Name: Network Analyzer Model Number(s): E5061B/E5063A Description and Non-volatile memory (Flash) for A51 DSP Module. This memory is for purpose product serial number and option. The system calibration data are also stored in the E5063A. (Any user data is not stored in these memory) Size 64 MB Memory clearing...
  • Page 10 Product Declassification and Security Product Name: Network Analyzer Model Number(s): E5061B/E5063A placed on it. This allows the E5061B/E5063A to still function properly in non-secured areas or for use when servicing. The E5061B-020 and all E5063A have a removable storage on the rear panel. Keysight has available a pre-configured storage for the E5061B/E5063A that must be purchased in order for this security method to work.
  • Page 11 Product Declassification and Security Product Name: Network Analyzer Model Number(s): E5061B/E5063A Storage Removal Procedure A 1. Clearly mark the spare hard drive as "Unsecured!" 2. If there is the file named *.lic under E:\License directory, copy it into your USB memory. This file contains software option information.
  • Page 12 Product Declassification and Security Product Name: Network Analyzer Model Number(s): E5061B/E5063A Storage Removal Procedure B 1. Loosen the 4 screws and remove the cover. (Keep the screws for re-assembly) 2. Loosen the 4 screws that hold bracket of the HDD/SSD. (Keep the screws for re- assembly) 3.
  • Page 13 Product Declassification and Security Product Name: Network Analyzer Model Number(s): E5061B/E5063A 6. Loosen the 4 screws (Keep the screws for re-assembly) 7. Remove the installed HDD/SDD Page 13 of 16...
  • Page 14 Product Declassification and Security Product Name: Network Analyzer Model Number(s): E5061B/E5063A Storage disk re-installation procedure: Once the E5061B/E5063A is returned to the secured area, follow the steps listed below. Any servicing of the E5061B/E5063A may include the regeneration of correction constants.
  • Page 15: User And Remote Interface Security

    Product Declassification and Security Product Name: Network Analyzer Model Number(s): E5061B/E5063A User and Remote Interface Security Measures Screen and Annotation Blanking The frequency-blanking feature is available. This function provides three security levels: “OFF” during normal operation; “Low” deletes frequency information from the display, but can be turned “OFF” by front panel operation;...
  • Page 16 Product Declassification and Security Product Name: Network Analyzer Model Number(s): E5061B/E5063A E5063A/E5061B Windows 7 & Windows 10 Refer to the following site. E5061B http://ena.support.keysight.com/e5061b/manuals/webhelp/eng/using_windows_xp/enablingdisabli ng_usb_storage.htm E5063A http://ena.support.keysight.com/e5063a/manuals/webhelp/eng/using_windows/enablingdisabling _usb_storage.htm Remote Access Interfaces The user is responsible for providing security for the I/O ports for remote access by controlling physical access to the I/O ports.