Execution
Conditions
Syntax
Parameters
Remarks
Example
Statements
NOTE
Syntax
Example
Statements
AZ
This command is available for the Keysight E5270B, and is used to enable or
disable the ADC zero function that is the function to cancel offset of the
high-resolution A/D converter. This function is especially effective for low voltage
measurements. Power on, *RST command, and device clear disable the function.
This command is effective for the high-resolution A/D converter, not effective for
the high-speed A/D converter.
This command is not available for the Keysight E5260 series.
AZ mode
mode :
Mode ON or OFF.
0: OFF. Disables the function. Initial setting.
1: ON. Enables the function.
Set the function to OFF in cases that the measurement speed is more important than
the measurement accuracy. This roughly halves the integration time.
OUTPUT @E5270;"AZ 0"
BC
The BC command clears the output data buffer that stores measurement data and
query command response data. This command does not change the measurement
settings.
Multi command statement is not allowed for this command.
BC
OUTPUT @E5270;"BC"