Error Messages
Error Codes
225
Send WSV, WSI, or WNX to get sync sweep data.
If you enable data output of the synchronous sweep source, do not
forget to set the synchronous sweep source by the WSV, WSI, or WNX
command. For data output, refer to
226
Set linear sweep for MM4 or MM5.
Only the linear sweep is available for the PWV or PWI command for
the pulsed sweep measurement (MM4) or the WV or WI command for
the staircase sweep with pulsed bias measurement (MM5).
227
Sweep measurement was aborted.
Sweep measurement was aborted by the automatic abort function or the
power compliance.
230
Pulse source must be set.
To perform the pulsed spot measurement (MM3), send the PV or PI
command to set the pulse source.
231
Compliance must be set correctly.
Compliance was not set or an incorrect compliance value was set in the
PV, PI, PWV, or PWI command. Set the correct compliance value
effective for the pulse output.
238
Too large pulse width (max. 2 s).
The maximum value of the pulse width is 2 s. And the available value
depends on the pulse period value. Refer to
239
Pulse width must be 0.5 ms or more.
Set the pulse width to 0.5 ms or more. Refer to
253
Program memory is full.
Maximum of 2000 programs or 40000 commands can be stored in the
program memory. Refer to
254
Invalid input for a memory program.
The GPIB GET command (TRIGGER statement in HP BASIC) and an
external trigger input are not allowed in a memory program (between
the ST and END commands).
255
Maximum nesting level is eight.
Nesting (one program calling another) of a memory program must be
eight levels or less.
5- 8