Enables channels
Disables channels
Sets filter ON/OFF
Sets series resistor ON/OFF
Sets pulse timing parameters
Sets auto abort function
Sets pulsed sweep source
Sets synchronous sweep source
Forces constant voltage
Forces constant current
Sets voltage measurement range
Sets current measurement range
Selects measurement mode
Sets SMU operation mode
Executes measurement
NOTE
Programming Examples
Pulsed Sweep Measurements
Pulsed Sweep Measurements
To perform pulsed sweep measurements, use the following commands.
Function
a. The WSV/WSI command must be entered after the PWV/PWI command.
Measurement channel performs measurement so that the pulse width and pulse
period are kept. The integration time is automatically set by the instrument, and you
cannot change. For the Keysight E5270B, note that the high-resolution ADC cannot
be used for the pulsed measurements. The AAD/AIT/AV/WT command settings are
ignored.
3- 28
Keysight E5260/E5270 Programming Guide, Edition 4
Command
CN
[chnum ... [,chnum] ... ]
CL
[chnum ... [,chnum] ... ]
[FL]
mode[,chnum ... [,chnum] ... ]
[SSR]
chnum[,mode]
PT
hold,width,period [,tdelay]
[WM]
abort[,post]
PWV
chnum,mode,range,base,start,
stop, step[,comp]
PWI
[WSV]
chnum,range,start,stop
a
[,comp[,Pcomp]]
[WSI]
DV, TDV
chnum,range,output
[,comp[,polarity[,crange]]]
DI, TDI
[RV]
chnum,range
[RI]
chnum,range
[RM]
chnum,mode[,rate]
MM
4,chnum
[CMM]
chnum,mode
XE
Parameters