Returning voltage and current data from the data buffer
The MEASure and FETch queries can also return all 4096 data values of the
instantaneous voltage and current buffers. These are as follows:
Regulatory-compliant measurement of quasi-stationary harmonics
In order to meet regulatory requirements of IEC-555 and other standards that
specify how quasi-stationary harmonics are to be measured, the ac source has a
command that alters both the output frequency control and the harmonic
measurement algorithms to meet these requirements. The command is
Keysight Models 6811C, 6812C, and 6813C Programming Guide
MEASure:ARRay:CURRent?
MEASure:ARRay:VOLTage?
SENSe:WINDow KBESsel | RECTangular
Programming Examples