Keysight Technologies E5260 Series Programming Manual page 263

Parametric measurement solutions
Hide thumbs Also See for E5260 Series:
Table of Contents

Advertisement

Syntax
Parameters
Example
Statements
LSTM
The LSTM command sets the timing parameters for the linear search measurement.
If you do not enter this command, all parameters are set to 0.
LSTM hold,delay
hold :
Hold time (in seconds) that is the wait time after starting the search
measurement and before starting the delay time for the first search
point. Numeric expression.
0 to 655.35 sec. 0.01 sec resolution.
delay :
Delay time (in seconds) that is the wait time after starting to force a step
output value and before starting a step measurement. Numeric
expression.
0 to 65.535 sec. 0.0001 sec resolution.
OUTPUT @E5270;"LSTM 5,0.1"
Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
4- 79