Keysight Technologies E5260 Series Programming Manual page 45

Parametric measurement solutions
Hide thumbs Also See for E5260 Series:
Table of Contents

Advertisement

A
B
C
Measurement or source output data type. One bit.
A
0
1
Data type. One bit.
B
0
1
Measurement or output range. Five bits.
C
01000 (8)
01001 (9)
01010 (10)
01011 (11)
01100 (12)
01101 (13)
01110 (14)
01111 (15)
10000 (16)
10001 (17)
10010 (18)
10011 (19)
10100 (20)
11111 (31)
Keysight E5260/E5270 Programming Guide, Edition 4
Explanation
Source output data.
Measurement data.
Explanation
Voltage data.
Current data.
Explanation
0.5 V or 1 pA range.
5 V or 10 pA range.
100 pA range.
2 V or 1 nA range.
20 V or 10 nA range.
40 V or 100 nA range.
100 V or 1 μA range.
200 V or 10 μA range.
100 μA range.
1 mA range.
10 mA range.
100 mA range.
1 A range for E5280B/E5290A HPSMU
or 200 mA range for E5291A MPSMU.
Invalid data is returned.
Programming Basics
Data Output Format