Example
Statements
If you set Pcomp, the maximum Vcomp value for the module is
allowed, regardless of the output range setting.
For the log sweep and without Pcomp, set the value available for the
minimum range that covers start and stop values.
Pcomp :
Power compliance (in W). Numeric expression. Resolution: 0.001 W. If
the Pcomp value is not entered, the power compliance is not set.
0.001 to 2, 0.001 to 4 for E5291A, or 0.001 to 20 for E5280B/E5290A
OUTPUT @E5270;"WSI 1,16,0,4E-5"
OUTPUT @E5270;"WSI 2,0,1E-3,1E-2,5,5E-2"
Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
4- 141