Syntax
Conditions after
Execution
Example
Statements
Command Reference
Command Reference
AB
The AB command aborts the present operation and subsequent command execution.
This command stops the operation now in progress, such as the measurement
execution, source setup changing, and so on. But this command does not change the
present condition. For example, if the E5260/E5270 just keeps to force the DC bias,
the AB command does not stop the DC bias output.
AB
The AB command sets the E5260/E5270 as listed in the following table.
Operation before AB
Staircase sweep measurement
Pulsed spot measurement
Pulsed sweep measurement
Staircase sweep with pulsed bias
measurement
Quasi-pulsed spot measurement
Linear search measurement
Binary search measurement
Multi channel sweep measurement
Self-test
Self-calibration
Wait state (PA/PAX/WS/WSX command)
Program execution (RU or DO command)
OUTPUT @E5270;"AB"
4- 18
Keysight E5260/E5270 Programming Guide, Edition 4
Setting after AB
Sets specified start value.
Sets specified base value.
Sets specified base value.
Sets specified start value and base
value.
Sets specified start value.
Sets specified start value.
Sets specified start value.
Sets specified start value.
Same as set by CL command.
Same as set by CL command.
Settings do not change.
Settings do not change.