The BGI command specifies the current monitor channel and its search parameters
for the binary search measurement. This command ignores the RI command setting.
This command setting is cleared by the BGV command.
BGI chnum,mode,condition,range,target
chnum :
Search monitor channel number. The value must be slot number where
the module has been installed. Integer expression. 1 to 8 (1 and 5 are
not available for HPSMU). See
mode,
condition : Search mode (0: limit mode or 1: repeat mode) and search stop
condition. The meaning of condition depends on the mode setting:
mode
0
1
range :
Measurement ranging type. Integer expression. The measurement range
will be set to the minimum range that covers the target value. For the
limited auto ranging, the instrument never uses the range less than the
specified range. See
target :
Search target current (in A). Numeric expression.
0 to ±0.1, 0 to ±0.2 for E5291A, or 0 to ±1 for E5280B/E5290A
OUTPUT @E5270;"BGI 1,0,1E-8,14,1E-6"
"BSM"
4- 28
Keysight E5260/E5270 Programming Guide, Edition 4
Table 4-1 on page
Limit value for the search target (target). The search stops
when the monitor data reaches target ± condition.
Numeric expression. Positive value. in A.
Setting resolution: range/20000. where range means the
measurement range actually used for the measurement.
Repeat count. The search stops when the repeat count of the
operation that changes the source output value is over the
specified value. Numeric expression. 1 to 16.
Table 4-3 on page
4-9.
condition
4-11.