Keysight Technologies E5260 Series Programming Manual page 311

Parametric measurement solutions
Hide thumbs Also See for E5260 Series:
Table of Contents

Advertisement

Execution
Conditions
Syntax
Parameters
Example
Statements
TV
The TV command performs the high speed spot measurement, and returns the
measurement data. The command starts a voltage measurement regardless of the
SMU operation mode, trigger mode (TM command), and measurement mode (MM
command).
CN command has been executed for the specified channel.
TV chnum[,range]
chnum :
Measurement channel number. The value must be slot number where
the module has been installed. Integer expression. 1 to 8 (1 and 5 are
not available for HPSMU). See
range :
Measurement range or ranging type. Integer expression. If you select
the fixed range, the instrument performs measurement by using the
specified range. For the auto or limited auto ranging, the measurement
range will be set to the minimum range that covers the measured
values. However, the instrument never uses the range less than the
specified range for the limited auto ranging. See
page
4-10.
If you do not specify the range parameter for current output channels,
the channel uses the minimum range that covers the compliance value.
The range parameter is meaningless for voltage output channels. The
measurement ranging type is always same as the output ranging type.
OUTPUT @E5270;"TV 1"
ENTER @E5270 USING "#,3X,12D,X";Vdata
Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
Table 4-1 on page
4-9.
Table 4-2 on
4- 127