Syntax
Parameters
Remarks
Example
Statements
See Also
Command Reference
Command Reference
TGPC
The TGPC command clears the trigger setting of the specified ports.
TGPC [port[,port...[,port]...]]
A maximum of 18 ports can be set. If no port is specified, the TGPC command
clears the setting of the all ports; Ext Trig In, Ext Trig Out, and digital I/O ports 1 to
16.
port :
Trigger port number. Integer expression. -1, -2, or 1 to 16.
-1: Ext Trig In terminal.
-2: Ext Trig Out terminal.
1 to 16: Port 1 to 16 of the digital I/O terminal.
The TGPC command sets the trigger ports as shown below.
Ext Trig In
Same as after TGP -1,1,2,0 command execution.
Ext Trig Out
Same as after TGP -2,2,2,0 command execution.
Digital I/O Ports No trigger function is available. The ERS? and ERC commands
are available for the port control.
This is not same as the condition set by the *RST command that sets the ports as
shown below.
Ext Trig In
Same as after TGP -1,1,2,1 command execution.
Ext Trig Out
Same as after TGP -2,2,2,1 command execution.
Digital I/O Ports No trigger function is available. The ERS? and ERC commands
are available for the port control.
OUTPUT @E5270;"TGPC -1,-2,1,2"
"TGP"