Table 1-4
Programming Basics
Data Output Format
Status for Data, Data_search, or Data_sense
E
000 (0)
No status error occurred.
001 (1)
Another channel reached its compliance setting.
010 (2)
This channel reached its compliance setting.
011 (3)
Measurement data is over the measurement range.
D will be the meaningless value 1111111111111111 (65535).
100 (4)
One or more channels are oscillating.
Or source output did not settle before measurement.
110 (6)
For linear or binary search measurement, the target value was not
found within the search range. Returns the source output value.
For quasi-pulsed spot measurement, the detection time was over the
limit (3 s for Short mode, 12 s for Long mode).
111 (7)
For linear or binary search measurement, the search measurement was
stopped. Returns the source output value. See status of Data_sense.
For quasi-pulsed spot measurement, output slew rate was too slow to
perform the settling detection.
reached the current compliance before the source output voltage
changed 10 V from the start voltage.
a. Make the wait time or delay time longer. Or make the current compli-
ance larger. For pulsed measurement, make the pulse width longer, or
make the pulse base value closer to the pulse peak value. For current
output by limited auto ranging, make the output range lower.
b. Make the current compliance or start voltage larger. Or set the detection
interval to Long. If this status occurs with the Long mode, perform the
spot measurement.
c. Make the current compliance larger. Or set the detection interval to
Long. If this status occurs with the Long mode, perform the spot mea-
surement or pulsed spot measurement.
d. Perform the pulsed spot measurement or spot measurement.
1- 36
Keysight E5260/E5270 Programming Guide, Edition 4
Explanation
b
c
Or quasi-pulsed source channel
d
a