NOTE
Measurement Modes
The Keysight E5260/E5270 provides the following measurement modes.
•
"Spot Measurements"
•
"Pulsed Spot Measurements"
•
"Staircase Sweep Measurements"
•
"Multi Channel Sweep Measurements"
•
"Pulsed Sweep Measurements"
•
"Staircase Sweep with Pulsed Bias Measurements"
•
"Quasi-Pulsed Spot Measurements"
•
"Binary Search Measurements"
•
"Linear Search Measurements"
About Search Measurements
The E5260/E5270 supports search measurement to find a point on an I-V curve
where a specified condition is satisfied. For example, it searches for a breakdown
voltage or threshold voltage at a specified current.
Search measurements are performed by one or two SMUs. For two SMUs, one is the
search channel, and the other is a sense channel. When one SMU is used, it serves as
both search and sense channel.
Basically, the search channel forces voltage or current until the search stop condition
is satisfied.
Keysight E5260/E5270 Programming Guide, Edition 4
Remote Mode Functions
Measurement Modes
2- 3