MS/MS operating modes

In SIR and MRM modes, neither quadrupole is scanned, so no spectrum
Note:
(intensity versus mass) is produced.
The following table shows the MS/MS operating modes.
MS/MS operating modes:
Operating mode
Product
(daughter) ion
spectrum
Precursor
(parent) ion
spectrum
MRM
Constant neutral
loss spectrum
ScanWave
daughter scan
RADAR™ is an additional mode with which you collect data from the MRM
and full scan MS modes simultaneously. RADAR mode can also acquire all
detectable ions in both positive and negative full scan MS.
1-14
Waters Xevo TQ-S Overview
MS1
Static (at
precursor mass)
Scanning
Static (at
precursor mass)
Scanning
(synchronized
with MS2)
Static (at
precursor mass)
Collision cell
Pass all masses
ScanWave
enabled
MS2
Scanning
Static (at product
mass)
Static (at product
mass)
Scanning
(synchronized
with MS1)
Scanning
(synchronized
with collision
cell)