Fitting The Stepwave Assembly To The Source Assembly - Waters Xevo TQ-S Operator's, Overview And Maintenance Manual

Mass spectrometry system
Table of Contents

Advertisement

Fitting the StepWave assembly to the source assembly

Required materials
Chemical-resistant, powder-free gloves
Seal breaker and locator tool
StepWave assembly removal and insertion tool
To fit the StepWave assembly to the source assembly:
Warning:
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Caution:
fitting it to the source assembly, use only the seal breaker and locator,
and StepWave assembly removal and insertion tools; do not use any
other tool.
1.
Pull the source enclosure release (located at the bottom, right-hand side)
outwards, and swing open the enclosure.
2.
Using both hands, grasp the source enclosure, and lift it vertically off the
two supporting studs on the source adaptor housing.
3.
Slide the first ion guide assembly end of the StepWave assembly into the
StepWave removal and insertion tool.
The StepWave assembly can only be inserted in the correct
Tip:
orientation in the StepWave removal and insertion tool.
The source components can be contaminated with
To avoid damage to the StepWave ion guide assembly when
Cleaning the StepWave ion guide assembly
4-75

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents