Running the mass spectrometer at high flow rates
The ACQUITY UPLC system runs at high flow rates. To optimize desolvation,
and thus sensitivity, run the ACQUITY Xevo TQ-S system at appropriate gas
flows and desolvation temperatures. IntelliStart software automatically sets
these parameters when you enter a flow rate, according to the following table.
Flow rate versus temperature and gas flow:
Flow rate
(mL/min)
0.000 to 0.020
0.021 to 0.100
0.101 to 0.500
>0.500
If you are using an APCI interface, IntelliStart software automatically sets
the parameters according to the following table.
Flow rate versus IonSABRE II probe temperature and gas flow:
Flow rate
(mL/min)
0.000 to 0.020
0.021 to 0.500
>0.500
Source
temperature (°C)
150
150
150
150
IonSABRE II
probe
temperature (°C)
400
500
0
Desolvation
temperature (°C)
200
300
500
0
Desolvation gas
flow (L/h)
800
1000
1000
Starting the mass spectrometer
Desolvation gas
flow (L/h)
800
800
1000
1000