Leaving the mass spectrometer ready for operation

Leave the mass spectrometer in Operate mode except in the following cases:
When performing routine maintenance
When changing the source
When leaving the mass spectrometer unused for a long period
In these instances, put the mass spectrometer in Standby mode (see the online
Help for details).

Emergency shutdown of the mass spectrometer

To shut down the mass spectrometer in an emergency:
Warning:
main power supply. To isolate the instrument, disconnect the power
cable from the back of the instrument.
Caution:
1.
Operate the power button on the front of the mass spectrometer.
2.
Disconnect the power cable from the rear panel.
2-10
Preparing the Mass Spectrometer for Operation
The instrument's power switch does not isolate it from the
Data can be lost during an emergency shutdown.