Cleaning or replacing the corona pin
Required materials
•
Chemical-resistant, powder-free gloves
•
Needle-nose pliers
•
HPLC-grade (or better) methanol
•
Lint-free tissue
•
Lapping film
•
Corona pin
To clean or replace the corona pin:
Warning:
with biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
great care while performing this procedure.
Warning:
Standby mode before commencing this procedure.
Warning:
the corona pin with care.
1.
Remove the corona pin from the source (see
2.
Replace the corona pin if it is deformed or otherwise damaged.
Otherwise clean its tip with the lapping film, and then wipe it clean with
a methanol-saturated tissue.
3.
Install the corona pin in the source (see
The probe and source components can be contaminated
The probe and source can be hot. To avoid burn injuries, take
To avoid electric shock, ensure that the instrument is in
The corona pin tip is sharp. To avoid puncture wounds, handle
page
4-15).
page
4-12).
Cleaning or replacing the corona pin
4-97