Measuring Short Compensation Data - Keysight 16196A Operation And Service Manual

Table of Contents

Advertisement

Operation
Performing Fixture Compensation

Measuring Short Compensation Data

Set the fixture in the short state using the short plate supplied.
Step 1. Remove the cap. Take out the open plate used to measure the open
compensation data.
Make sure there is no dirt or other foreign matter in the DUT insertion hole.
An exclusive type of short plate is supplied with each model. Do not use a
short plate from a different model.
Step 2. Place the short plate on the insulator assembly with tweezers. Place the
rod-shaped protrusion of the short plate downward, and insert it into the DUT
insertion hole.
Handle the short plate with Tweezers. If dirt, etc. gets on it, measuring
precision and repeatability may be adversely affected.
Figure 3-5
Setting the Short State Using the Short Plate
Step 3. Fit the cap in place with the mark toward the front, and turn it to the right
until it is locked.
Keysight 16196A/B/C/D Parallel Electrode SMD Test Fixture
27

Advertisement

Table of Contents
loading

This manual is also suitable for:

16196b16196c16196d

Table of Contents