Summary of Contents for Keysight Technologies 11644A Series
Page 1
Keysight Technologies 11644A X, P, and K Waveguide Calibration Kits This manual applies to 11644A series calibration kits with serial number prefix 3032A. User’s and Service Guide...
Page 4
CONTROL. Assistance Product maintenance agreements and other customer assistance agreements are available for Keysight products. For any assistance, contact Keysight Technologies. Refer to page 6-4 for a list of Keysight contacts. Printing Copies of Documentation from the Web To print copies of documentation from the Web, download the PDF file from the Keysight web site: •...
Use the Contents List in the shipping container to verify the completeness of your shipment. Although this list is the most accurate, you can also use the illustrations in Chapter 7 to verify the items in your shipment. If your shipment is not complete, contact Keysight Technologies - refer to“Contacting Keysight” on page 6-4.
Calibration Kit History This manual applies to any Keysight 11644A series (X, P, or K) waveguide calibration kits whose serial number prefix is 3032A and above. If your calibration kit has a different serial number prefix, refer to the next section for information on how this manual applies.
General Information Recording the Device Serial Numbers Recording the Device Serial Numbers In addition to the kit serial number, the devices in the kit are individually serialized (serial numbers are labeled onto the body of each device). Record these serial numbers in the appropriate table. Recording the serial numbers will prevent confusing the devices in this kit with similar devices from other kits.
General Information Preventive Maintenance Preventive Maintenance The best techniques for maintaining the integrity of the devices in the kit include: • routine visual inspection • routine cleaning • proper gaging • proper connection techniques All of these are described in Chapter 4.
Page 12
General Information Preventive Maintenance 1- 6 X, P, and K 11644A...
Specifications Environmental Requirements Environmental Requirements Table 2-1 Environmental Requirements Parameter Limits Temperature +20 C to +26 C Operating Storage 40 C to +75 C Error corrected range 1 C of measurement calibration temperature Relative humidity Type tested, 0% to 95% at 40 C, non-condensing 1.
Specifications Mechanical Characteristics Mechanical Characteristics Table 2-2 Table 2-3 list the typical characteristics of the adapters and 1/4 wavelength shims in your kit. These are not specifications, but are included as additional information. Table 2-2 Mechanical Characteristics of Adapters Frequency band Insertion Center Conductor Pin Recession...
Specifications Electrical Specifications Electrical Specifications Table 2-4 through Table 2-5 list the electrical specifications of the terminations, ¼ wavelength shims, and standard sections in your kit. Table 2-4 Electrical Characteristics and Specifications Device Frequency Termination Specification X11644A WR-90 8.2 to 12.4 GHz return loss 42d...
Application Notes and Technical Papers hyperlink. Certification Keysight Technologies certifies that this product met its published specifications at the time of shipment from the factory. Keysight further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (NIST) to the extent allowed by the institute’s calibration facility, and to the calibration facilities of other International Standards Organization members.
User Information Calibration Devices and Their Use Calibration Devices and Their Use The P, K, and X11644A waveguide calibration kits contain termination load, standard sections, shim, and short. For measurement convenience, these kits contain either 7 mm coax-to-waveguide adapters for P-band and X-band, or 3.5 mm coax-to-waveguide adapters for K-band.
User Information Measurement Applications Measurement Applications NOTE For your convenience, two different lengths of screws are provided in this kit. While you can use the long screws for any connection, the shorter screws provide a faster connection for two-flange connections. Isolation In most cases, select the softkey on your network analyzer.
User Information Measurement Applications Load Connect the termination to the appropriate test port, as shown in Figure 3-3. Figure 3-3 Test Port and Termination Offset Load Connect the shim and the termination to the appropriate test port, as shown in Figure 3-4.
User Information Measurement Applications Offset Short Connect the shim and the short to the appropriate test port, as shown in Figure 3-5. Figure 3-5 Connect the Shim, Short, and Test Port Reflect Connect the short to the appropriate test port, as shown in Figure 3-6.
User Information Measurement Applications Thru No device is required for this. Connect port 1 to port 2, as shown in Figure 3-7. Figure 3-7 Port 1 and Port 2 ¼ Wavelength Load “Offset Load” ¼ Wavelength Short “Offset Short” 3- 6 X, P, and K 11644A...
User Information Changing the ¼ Wavelength Shim Calibration Definition Changing the ¼ Wavelength Shim Calibration Definition The calibration kit definition data provided with this kit has a nominal value for the ¼ wavelength shim offset delay. You may use the nominal value provided, or measure the exact thickness of the shim and use that value to calculate its exact offset delay.
Page 26
User Information Changing the ¼ Wavelength Shim Calibration Definition 3- 8 X, P, and K 11644A...
Use, Maintenance, and Care of the Devices Electrostatic Discharge Electrostatic Discharge Protection against electrostatic discharge (ESD) is essential while connecting, inspecting, or cleaning connectors attached to a static-sensitive circuit (such as those found in test sets). Static electricity can build up on your body and can easily damage sensitive internal circuit elements when discharged.
Use, Maintenance, and Care of the Devices Visual Inspection Visual Inspection Visual inspection and, if necessary, cleaning should be done every time a connection is made. Inspect mating surfaces for dirt, dust, foreign particles, or scratches, which can degrade device performance. A damaged mating surface can damage any good surface connected to it.
Use, Maintenance, and Care of the Devices Cleaning the Mating Plane Surfaces Cleaning the Mating Plane Surfaces 1. Use Compressed Air or Nitrogen WARNING Always use protective eyewear when using compressed air or nitrogen. Use compressed air (or nitrogen) to loosen particles on the mating plane surfaces. Clean air cannot damage a device or leave particles or residues behind.
Use, Maintenance, and Care of the Devices Connections Connections Good connections require a skilled operator. Slight errors in operator technique can have a significant effect on measurements and measurement uncertainties. The most common cause of measurement error is poor connections. The following procedures illustrate how to make good connections.
Use, Maintenance, and Care of the Devices Connections Aligning Two Precision Flanges 1. Place the slip pins in the top and bottom holes of one flange, as shown in Figure 4-3. Figure 4-3 Inserting Slip Pins 2. Using the pins as guides, carefully align the flanges and insert two screws in the diagonal corner holes, as shown in Figure 4-4.
Use, Maintenance, and Care of the Devices Connections Aligning a Precision and a Non-Precision Flange 1. Place an alignment pin (with head) in the corner hole of one flange. Place a second alignment pin in the diagonal corner hole of the second flange, as shown in Figure 4-5.
Use, Maintenance, and Care of the Devices Connections Tightening a Flange Connection NOTE The best connection has symmetrical pressure applied as you gradually tighten the screws. 1. In an “X” pattern (for equal compression), tighten all four screws using a hex ball driver. Do not over-tighten.
Use, Maintenance, and Care of the Devices Handling and Storage Handling and Storage • Install the protective end caps and store the calibration devices in the foam-lined storage case when not in use. • Never store devices loose in a box, or in a desk or bench drawer. This is the most common cause of device damage during storage.
Page 36
Use, Maintenance, and Care of the Devices Handling and Storage 4-10 X, P, and K 11644A...
Introduction Introduction The performance of your calibration kit can only be verified by returning the kit to Keysight Technologies for recertification. The equipment and calibration standards required to verify the specifications of the limits of the devices in the kit have been specially manufactured and are not commercially available.
A list of NIST traceable numbers may be purchased upon request to be included in the calibration report. Keysight Technologies offers a Standard calibration for the recertification of the kit. For more information, contact Keysight Technologies. Refer to “Contacting Keysight” on page 6-4 for a list of offices.
Performance Verification Performance Test Performance Test Termination Return Loss Measurement For your convenience the following procedures have been included in this kit. Use this test, as needed, to check the performance of the terminations in this kit. Figure 5-1 Return Loss Test Setup 1.
Performance Verification Performance Test 4. Set the averaging factor to 512. 5. At the adapter test port, perform a 2–port TRL calibration. 6. Turn on the calibration. 7. Connect the termination you wish to test to port 1. 8. Measure the return loss (S ) of the load.
Performance Verification System Operation Checks System Operation Checks Use the following procedures and the standard section in this kit to verify the operation of your calibrated system. NOTE If you are using a different analyzer, refer to its documentation for specific measurement instructions.
Page 43
Performance Verification System Operation Checks 7. To set the display: a. Press b. Press RESPONSE [SCALE] [.1] [x1] c. Select [REF POSN] [10] [x1] d. Select [REF VALUE] [0] [x1] e. Press MEASUREMENT [RESTART] 8. After one complete measurement sweep, the displayed trace should look similar to the typical trace shown in Figure 5-3.
Performance Verification System Operation Checks The S and S of a Thru Measurement 1. Verify that the calibration is still on. 2. Verify that the averaging factor is set to 512. 3. Connect the standard section to ports 1 and 2 to form a thru connection. 4.
Page 45
Performance Verification System Operation Checks The S and S of a Thru Measurement 1. Verify that the calibration is still on. 2. Verify that the averaging factor is set to 512. 3. Connect the standard section to ports 1 and 2 to form a thru connection. 4.
Performance Verification System Operation Checks In Case of Failure If a system check fails, recalibrate the system and repeat the entire procedure (all three measurements). Remember that poor connections are the most common cause of measurement errors. Also, both the hardware state and the instrument state must be correct.
Page 47
Performance Verification System Operation Checks Table 5-1 Using a TRL Calibration X11644A WR-90 (X-band) Test Measured Acceptable Pass Value Value (Yes/No) , shorted 0.15 dB , shorted 0.15 dB , thru 45 dB , thru 45 dB , thru 0.017 dB , thru 0.017 dB Table 5-2 Using a TRL Calibration P11644A WR-62 (P-band)
Performance Verification System Operation Checks 872x System Operation Check 1. Short the standard section and perform an S and an S measurement to verify that the ripple on the reflection of the shorted waveguide section does not exceed the limits listed in the tables on page 5-15.
Page 49
Performance Verification System Operation Checks 9. Record the maximum peak-to-peak value in the appropriate table on page 5-15. 10. Turn off the delta marker mode. 11. Repeat this procedure from step 6, and make an S measurement on the standard section. Record the measured value in the appropriate table on page 5-15.
Performance Verification System Operation Checks 8. Repeat this procedure from step 3, and make an S measurement. Record the measured value in the appropriate table on page 5-15. 9. If you use the delta marker mode, don’t forget to turn it off. In Case of Failure If a system check fails, recalibrate the system and repeat the entire procedure (all three measurements).
Page 51
Performance Verification System Operation Checks Table 5-4 Using a Full 2-Port Calibration X11644A WR-90 (X-band) Test Measured Acceptable Pass Value Value (Yes/No) , shorted 0.15 dB , shorted 0.15 dB , thru 40 dB , thru 40 dB , thru 0.06 dB , thru 0.06 dB...
Page 52
Performance Verification System Operation Checks 5-16 X, P, and K 11644A...
Troubleshooting Troubleshooting Process Troubleshooting Process This manual contains limited information about network analyzer system operation. For complete information, refer to the instrument documentation. If you suspect a bad calibration, or if your network analyzer does not pass performance verification, follow the steps in Figure 6-1 Figure 6-1...
6-4. Returning a Kit or Device to Keysight If your kit or device requires service, contact Keysight Technologies for information on where to send it - “Contacting Keysight” on page 6-4. Include a service tag (located near the end of this manual) on which you provide the following information: •...
Troubleshooting Contacting Keysight Contacting Keysight Assistance with test and measurement needs and information on finding a local Keysight office are available on the Web at: www.keysight.com/find/assist If you do not have access to the Internet, please contact your Keysight field engineer. NOTE In any correspondence or telephone conversation, refer to the Keysight product by its model number and full serial number.
To order a listed part, note the description, part number, and the quantity desired. Telephone or send your order to Keysight Technologies. Refer to “Contacting Keysight” on page 6-4 for further information.
Page 60
Replaceable Parts Introduction Table 7-1 Replaceable Parts for the X11644A WR-90 Description Keysight Per Kit Part Number Calibration Devices Termination 00910-60003 Standard section 00896-60008 Short 11644-20018 ¼ Wavelength shim 11644-20021 7 mm coaxtowaveguide adapter X281C Option 006 Hardware Alignment pin 11644-20024 Slip pin 11644-20025...
Page 61
Replaceable Parts Introduction Figure 7-1 X-Band Component Identification Sheet X, P, and K 11644A...
Page 62
Replaceable Parts Introduction Table 7-2 Replaceable Parts for the P11644A WR-62 Keysight Description Per Kit Part Number Calibration Devices Termination 00910-60002 Standard section 00896-60007 Short 11644-20017 ¼ Wavelength shim 11644-20020 7 mm coaxtowaveguide adapter P281C Option 006 Hardware Alignment pin 11644-20023 Slip pin 11644-20025...
Page 63
Replaceable Parts Introduction Figure 7-2 P-Band Component Identification Sheet X, P, and K 11644A...
Page 64
Replaceable Parts Introduction Table 7-3 Replaceable Parts for the K11644A WR-42 Description Keysight Per Kit Part Number Calibration Devices Termination 00910-60001 Standard section 00896-60006 Short 11644-20016 ¼ Wavelength shim 11644-20019 3.5 mm coaxtowaveguide adapter (m) 00281-60001 3.5 mm coaxtowaveguide adapter (f) K281C Option 006 Hardware Alignment pin...
Page 65
Replaceable Parts Introduction Figure 7-3 K-Band Component Identification Sheet X, P, and K 11644A...
Page 66
Replaceable Parts Introduction Table 7-4 Items Not Included in the Calibration Kit Description Keysight Part Number ESD Protection Devices Grounding wrist strap 9300-1367 5 ft grounding cord for wrist strap 9300-0980 2 ft by 4 ft conductive table mat with 15 ft grounding wire 9300-0797 ESD heel strap 9300-1308...
Standard Definitions Class Assignments and Standard Definitions Values are Available on the Web Class Assignments and Standard Definitions Values are Available on the Class assignments and standard definitions may change as more accurate model and calibration methods are developed. You can download the most accurate class assignments and standard definitions from Keysight’s Calibration Kit Definitions Web page at http://na.support.keysight.com/pna/caldefs/stddefs.html.
Page 69
See calibration flowchart, troubleshooting when to perform definitions frequency of calibration calibration applications contacting Keysight Technologies frequency specification 1/4 load 1/4 short contents line calibration kit load drawing of general information...
Page 70
7-10 deviation from nominal phase Keysight Technologies oxygen device application note electrical contacting environmental part numbers frequency contents included in kit...
Page 71
Index uncertainty, measurement verification device performance temperature visual inspection warranty, documentation waveguide, connecting wavelength shim wavelength shim, calibration definition when to calibrate wrench part number wrist strap for ESD protection part number 7-10 X, P, and K 11644A Index-3...
Need help?
Do you have a question about the 11644A Series and is the answer not in the manual?
Questions and answers