Sensors

Liquid level sensors are built into the caps of all bottles to avoid
overflow of the waste bottle and to warn the user when the buffer
bottles are almost empty.
The liquid level sensors can be enabled or disabled in the Sensors
window.
Caution If the sensors are disabled, there will be no warning prior to the
buffer bottle being empty or the waste bottle full. Failure may result in a
weakened or faulty wash result, liquid spillage, or biohazard.
Select the Sensors row in the Settings menu and press the
The Sensors window has the following parameters:
Buffer A sensor – Disable or enable the buffer A sensor. The default
is Enabled.
Buffer B sensor – Disable or enable the buffer B sensor. The default
is Enabled.
Buffer Rinse sensor – Disable or enable the buffer Rinse sensor.
The default is Enabled.
Buffer Waste sensor – Disable or enable the buffer Waste sensor.
The default is Enabled.
Plate sensor – Disable or enable the Plate sensor. The default is
Enabled. When enabled, the instrument checks whether there is a
plate on the plate carrier at the beginning of the protocol run.
Caution If the plate sensor is disabled, there will not be a warning
indicating a missing plate. A missing plate may result in spillage.
Volume (ml) – The autoprime volume ranges from 1 ml to
30 ml in 1 ml increments. The default is 5 ml.
Time (hh:mm) – The time ranges from 0:00 (= Off) to 9:00
in increments of 15 min. The default is Off.