Menus, Tabs and Parameters
Main menu
80
Thermo Scientific Wellwash Versa User Manual
You can stop a protocol momentarily and then start the protocol again.
Pause
The Pause step has the following parameters:
Buzzer – Set the buzzer on (= Yes) or off (= No). There is a beep
●
when the Pause step starts. The default is No.
Comment – You can add text, for example, instructions. There is
●
space for 128 characters.
To end the Pause step, press the
execution. A three-second alarm is sounded.
Layout
The Layout step is used to predefine the strip selection for a protocol.
This step determines the area on which the next steps are processed.
The layout is landscape when a 1x8, 2x8, or 1x16 wash head is
selected and portrait for a 1x12 or 2x12 wash head. The strips are
numbered 1–12 if an nx8 wash head is used and marked A–H for an
nx12 wash head. The strips are numbered 1–24 if a 1x16 wash head is
used.
If there is no Layout step in the protocol, the processed area is according
to the Main > Layout row selection.
Note Only the strips that are selected to the Layout step are saved with
the protocol. Strips selected using the Layout row are not saved.
button to continue protocol
OK
▲
Thermo Fisher Scientific