Menus, Tabs and Parameters
Maintenance menu
Clean wash head
90
Thermo Scientific Wellwash Versa User Manual
The procedure cleans the wash head according to a set of actions. Carry
out this procedure if the wash heads are dirty or clogged.
The instrument fills the priming vessel with a selected liquid and soaks
the wash head for a selected period of time.
Note Use a suitable cleaning agent and repeat the Clean procedure with
deionized distilled water.
1. Select the Clean wash head row and press the
2. Change the wash buffer to cleaning agent or deionized distilled
water. Press the
OK
▲
button.
button.
OK