Menus, Tabs and Parameters
Maintenance menu

Soak wash head

92
Thermo Scientific Wellwash Versa User Manual
The procedure soaks the wash head according to a set of actions. The
wash head is soaked in liquid (standby) to prevent clogging. The
instrument fills the priming vessel with a selected liquid and leaves the
wash head immersed in the solution. This procedure can be used for a
shorter period of time between washes using a wash buffer or to soak for
an extended period, for example, overnight, using deionized distilled
water (Rinse channel).
1. Select the Soak wash head row using the Down arrow key and
press the
button.
OK
2. Change the wash buffer to cleaning agent or deionized distilled
water if needed. Press the
3. Select the priming channel using the Down arrow key and/or press
the
button.
OK
Priming, preparing the soak and soaking take place.
button.
OK