Safety Specifications; In Conformity With The Requirements - Thermo Scientific Wellwash Versa User Manual

Hide thumbs Also See for Wellwash Versa:
Table of Contents

Advertisement

specifications
In conformity with
the requirements
Thermo Fisher Scientific
Safety
This section describes the safety specifications for the Wellwash Versa
instrument.
Wellwash Versa bears the following markings:
Type 888
100–240 Vac, 50/60 Hz, 100 VA max., Class I
CE marking
cTÜVus
The safety specifications are also met under the following environmental
conditions in addition to or in excess of those stated in the operating conditions:
Altitude
Temperature
Humidity
Mains supply fluctuations
Installation category
(overvoltage category)
Pollution degree
Note 1) The installation category (overvoltage category) defines the level
of transient overvoltage which the instrument is designed to withstand
safely. It depends on the nature of the electricity supply and its
overvoltage protection means. For example, in CAT II which is the
category used for instruments in installations supplied from a supply
comparable to public mains, such as hospital and research laboratories
and most industrial laboratories, the expected transient overvoltage is
2500 V for a 230 V supply and 1500 V for a 120 V supply.
2) The pollution degree describes the amount of conductive pollution
present in the operating environment. Pollution degree 2 assumes that
normally only nonconductive pollution, such as dust, occurs with the
exception of occasional conductivity caused by condensation.
Up to 2000 m
+5°C to +40°C
Maximum relative humidity 80% for temperatures up to
31°C decreasing linearly to 50% relative humidity at 40°C
± 10% from nominal
II according to IEC 60664-1 (see Note 1)
2 according to IEC 60664-1 (see Note 2)
Thermo Scientific Wellwash Versa User Manual
Technical Specifications

Safety specifications

131