11. Press the
1 key to continue calibration.
F
OR
Press the
2 key to cancel calibration.
F
OR
Press the
3 key to repeat calibration when the software informs
F
that the Delta% or accuracy are not acceptable. (The accuracy
must be 2.5–3.5 x that of 100 μl.)
12. Calibration is now complete. Press the
calibration.
Note Calibration for the wash head is saved and it becomes the active
calibration.
13. Press the
button when the procedure has finished.
OK
14. Verify the new performance by dispensing a plate, for example, with
300 μl. Tare weigh the plate, dispense and measure the weight to
determine the actual dispensed volume.
Calculation method: Weight [mg] / 96 = Volume/well [μl]