Filling And Emptying Liquid Bottles; Starting Ready-Made Protocols - Thermo Scientific Wellwash Versa User Manual

Hide thumbs Also See for Wellwash Versa:
Table of Contents

Advertisement

Running Protocols

Filling and emptying liquid bottles

Filling and
emptying liquid
Starting ready-
made protocols
44
Thermo Scientific Wellwash Versa User Manual
7. Press the
8. The microplate is processed according to the predefined protocol.
9. Check that the liquid channels work properly during dispensing and
aspiration.
10. If you need to abort the run, press the
11. Remove the plate after the protocol has ended.
12. Rinse the instrument. Refer to "Rinsing the liquid system" on page
51.
Note If you use a partial strip plate, make sure that your column/strip
selection matches the physical strips on the plate.
Note Do not remove the plate before the end of the protocol run.
Note It is recommended to shut down the Wellwash Versa at the end of
the day. Refer to "Shutting down" on page 109.
1. Disconnect the liquid bottles and fill or empty them if needed.
Refer to "Liquid level detection" on page 42.
bottles
2. Unscrew the liquid bottle cap and fill with a suitable buffer solution.
3. Replace the liquid bottle cap and reinstall the liquid or waste bottle.
Refer to "Liquid bottles and channels" on page 22.
4. Always reprime the system before running a protocol. Refer to
"Priming the system" on page 47.
You can select a protocol using the quick keys or from the protocol list
and start it.
button.
START
button during the run.
STOP