Installation

Settings after installation

Settings after
installation
32
Thermo Scientific Wellwash Versa User Manual
icon should be
red
when the waste bottle is in an upright position. For more
information, refer to "Liquid level detection" on page 42.
Fill the buffer bottles completely with liquid.
Prime all channels of the instrument to ensure that priming works;
the liquid tubing is filled with liquid and is aspirated to the waste
bottle. Check for possible air leaks in the liquid system. Lift the
buffer bottle about 20 cm and ensure that the wash head tips are not
dripping. For more information, refer to "Priming the system" on
page 47.
Check that the wash head configuration in the Settings menu
corresponds to the installed wash head. The information is used to
check for protocol wash head mismatches. Refer to "Wash head
configuration" on page 84.
Run a protocol, such as one of the demo protocols supplied with the
instrument. The protocol should have an aspirate, dispense, shake,
and soak function to ensure an adequate sample of functions to test
the instrument's proper operation. Check that the liquid channels
and wash head tips work properly during dispensing and aspiration.
Warning Clogged tips may cause a faulty washing performance.
To change the wash head type, the preferred user interface language or
the date/time, refer to "Settings menu" on page 81.