Menus, Tabs and Parameters
Maintenance menu

Starting the calibration

96
Thermo Scientific Wellwash Versa User Manual
the accuracy % for reference use only. The instrument uses linear
regression to calculate the dispense valve open time versus volume.
Note Ensure that the wash head type corresponds to the wash head
setting in the Settings menu.
Note Before running the calibration, make sure that the liquid system
operates correctly, for example, that the liquid filters and wash head tips
are clean.
Select the Calibration row in the Maintenance menu and press the
button.
OK