Running Protocols

Strip selection

Strip selection
48
Thermo Scientific Wellwash Versa User Manual
Figure 5–34. Dead volumes of the liquid tubing
Caution Prime the liquid system with the liquid to be used (wash
buffer). Ensure that the dispensing pump is not run for longer than a
few minutes without liquid, otherwise it may be damaged. Check for
possible air leaks.
Note If the instrument has not been used for a longer time, you might
need to perform the Boost prime option. Refer to "Service" on page
102.
Note In case of two-buffer protocols, both the buffer channels have to
be primed before the protocol is started. If channels are switched within
the same protocol, add a Prime step to the protocol to ensure that liquid
in the tubing is replaced completely (a volume of at least 15 ml is
required).
Strip selection can be carried out in two ways:
1. To select the columns/strips of the plate, use the number keys.
OR
2. Press the OK button on the Layout row in the Main menu.
Instrument to wash head
Bottle to instrument
8ml
Bottle to wash head
20ml
12ml