When you set up your Wellwash Versa, avoid sites with excess dust,
vibrations, strong magnetic fields, direct sunlight, draft, excessive
moisture, or large temperature fluctuations. Make sure that:
The working area is flat, dry, clean and vibration-proof, and leave
additional room for cables, covers, and so on.
There is at least 10 cm around the instrument on the laboratory
bench of free space for ventilation.
There is sufficient room behind the instrument to enable
disconnecting the device.
The ambient air is clean and free of corrosive vapors, smoke, and
dust.
The ambient temperature range is between +10°C (50°F) and +40°C
(104°F).
The humidity is low so that condensation does not occur (relative
humidity is between 10% and 80%, non-condensing).
Caution Do not operate the instrument in an environment where
potentially damaging liquids or gases are present.
Setups
This section describes the procedures that must be carried out before
instrument operation.
Warning All parts of the instrument that come into contact with
potentially infectious materials must be treated as potentially infectious
areas.
It is advisable to adhere to applicable safety precautions, such as the
wearing of disposable powder-free gloves, safety glasses, and protective
clothing, to avoid potential infectious disease contamination when
performing cleaning procedures and also when making adjustments to
the instrument.
Caution Leave the instrument to stand for at least three hours before
installing and switching it on, so there is no possibility of condensation
causing a short circuit.
The instrument comes with one transport lock.
Caution Make sure that the transport lock has been removed and that
the priming vessel, the wash heads, and the liquid bottles with tubes and
liquid level detectors have been installed before you put the instrument
into operation.