Figure 2–8. Priming vessel installation
Warning The priming vessel may be contaminated after the instrument
has been used.
The wash heads are packed separately and must be installed before use.
Keep the wash head package for storage purposes. If the wash head is
head(s)
not used, place it first into the enclosed plastic bag and then into the
original package. The wash heads supplied with the instrument are
calibrated at the factory. The wash head box includes an instrument
serial number according to which it has been calibrated at the factory.
Optional wash heads should be calibrated by the user prior to use.
Warning The wash heads may be contaminated after the instrument
has been used.
Warning Failing to connect the dispensing or aspiration tube may cause
spillage.
Caution Only use wash heads that have an identification label.
Note Always wear disposable powder-free gloves when handling the
wash heads.
Note Do not remove the lot number sticker on the wash head.
1. The instrument is supplied with a 1x8, 2x8, 1x12, or 2x12 wash
head.
2. Fit the supplied wash heads to the tubing according to Figure 2–9
through Figure 2–11. Note the difference in sizes.
The larger aspiration (waste) tube is connected to the
bottom hole.