Installation
Setups
26
Thermo Scientific Wellwash Versa User Manual
Figure 2–20. Liquid bottles connected to the instrument
Warning Make sure that the liquid level of the waste bottle is always
kept below the maximum level indicated on the bottle to avoid potential
overflow, as the contents of the waste bottle is potentially infectious.
Warning The contents of the waste bottle is potentially infectious, so it
is important to wear protective clothing, such as disposable gloves, a
laboratory coat, and safety glasses, when emptying or handling the waste
bottle.
Caution Always ensure that the liquid bottles are attached properly.
Always connect the bottle cap to the correct bottle and to the correct
connector on the left side of the instrument. Otherwise the wash
performance can be seriously affected.
Caution Ensure that the intake tube reaches fully down to the bottom
of the liquid bottle so that no red tubing is visible.
Caution Check regularly that the liquid filters in the intake tubes are
clean. If not, replace or clean them.
Caution To ensure proper function, do not disconnect the sensor cables
from the instrument during operation.