Maintenance menu
to
100
Thermo Scientific Wellwash Versa User Manual
8. Weigh the dispensed plate and enter the mass of the dispensed
liquid in grams using the number keys.
9. Press the
1 key to continue calibration.
F
OR
Press the
2 key to cancel calibration.
F
OR
Press the
3 key to repeat calibration when the software informs
F
that the Delta% or accuracy are not acceptable. (The accuracy of
the first plate must be ± 50% and that of the second plate must be
± 5%.)
10. Continue dispensing and weighing the third and fourth plate in
the same way.