Maintenance

Refitting the transport lock

Refitting the
transport lock
124
Thermo Scientific Wellwash Versa User Manual
7. Place the instrument in a large plastic bag.
8. Place a cloth soaked in the prepared decontaminant solution into
the bag. Ensure that the cloth does not come into contact with the
instrument.
9. Seal the bag firmly and leave the instrument in the bag for at least
24 hours.
10. Remove the instrument from the bag.
11. After decontamination, clean the instrument using a mild detergent.
12. After performing this decontamination procedure, enclose a signed
and dated "Certificate of Decontamination" (see Appendix B) both
inside the transport package and attached to the outside of the
package.
When you relocate the instrument or ship it for service, make sure that
you refit the transport lock.
1. Switch off the instrument.
2. Align the wash head arm and the plate carrier in order to refit the
transport lock screw (Figure 11–47).
Figure 11–47. Aligning the plate carrier and wash head arm
3. Screw on the transport lock screw and the transport lock tag with
the supplied Allen key (Figure 11–48).