Installation
Setups
Caution Do not cover the venting holes.
▲
Caution If foaming occurs in the waste bottle, refer to "Foaming" on
page 27.
▲
Caution Do not limit, change or remove the float position on the liquid
level sensor.
▲
Float
Figure 2–16. Correct position of the float on the buffer bottle sensor
Smaller bottles can also be used with an optional bottle holder
(Cat. No. N10820).
Caution If a smaller size buffer bottle is used with the bottle stand, the
remaining volume in the bottle after the liquid level sensor indicates
empty may be smaller than the nominal 290 ml.
▲
Caution Before disconnecting the tubing, prime the liquid system with
air to avoid liquid spillage.
▲
Figure 2–17 shows the connections of the liquid level sensor cabling and
Connection diagram
liquid bottle tubing.
Thermo Fisher Scientific
Thermo Scientific Wellwash Versa User Manual
23