Thermo Scientific Wellwash Versa User Manual page 125

Hide thumbs Also See for Wellwash Versa:
Table of Contents

Advertisement

Thermo Fisher Scientific
Warning The decontamination procedure should be performed by
authorized trained personnel in a well-ventilated room wearing
disposable gloves, protective glasses, and clothing.
Decontamination should be performed in accordance with normal
laboratory procedures. Any decontamination instructions provided with
the reagents used should be followed.
It is strongly recommended to perform the complete decontamination
procedure before relocating the instrument from one laboratory to
another.
Example of decontaminants:
Ethanol 70%
Virkon™ solution 1–3%
Glutaraldehyde solution 4%
Chloramine T
Microcide SQ™ 1:64
Decon™ 90 min. 4%
Caution If local or laboratory regulations prescribe regular
decontamination, it is not advisable to use formaldehyde, since even
small traces of formaldehyde negatively affect the enzyme being used in
EIA tests, resulting in inconsistent test results.
1. Prepare the decontaminant.
2. Run the Maintenance > Cleaning/Disinfection procedure using
the prepared decontaminant. Refer to "Cleaning/Disinfection
procedure" on page 87.
3. Autoclave the liquid bottles, tubes and liquid connectors if needed
(Figure 11–44). However, the liquid level sensors or the liquid
filters cannot be autoclaved (Figure 11–44).
4. Empty the plate carrier. You must wear disposable powder-free
gloves.
5. Switch off the power and disconnect the power supply cable.
6. Disinfect the outside of the instrument using a cloth dampened
with 70% ethanol.
Decontamination procedure
Thermo Scientific Wellwash Versa User Manual
Maintenance
123

Advertisement

Table of Contents
loading

Table of Contents